Description
- Description: fully automatic-random access, continuous loading, benchtop analyzer for Clinical Chemistry and Immunoturbidimetric assays.
- Assay type: End Point, Kinetic, Differential, Bichromatic and Immunoturbidimetric,1-3 reagent assays and all user programmable.
- Test entry mode: selective – profiles – batch – stat On line reagents: 1 cooled trays, 30 reagent positions and 60 samples positions including standard, control and stat.
- Primary tubes size: 16x100mm
- Stat samples processing: allowed at any time
- Throughput: 280 test/hr without ISE
- Reading system: direct photometry
- Minimum reaction volume: 180 ul
- Optics system Photometer: multi-wavelength optics
- Wavelength: 8 narrow band interferential filters
- Light source: halogen lamp 12V/50W
- Optical path: 8 mm.
- Linear range: 0.001 – 2.500
- Abs Resolution: 0,001 Abs
- Dispensingarm: one sample/two reagent arms with: – capacitive liquid level sensing – reagent pre-warming at 37C – automatic probe washin
Diluter: dilutions are made without using traditional syringes with the following specifications: – sample volume: 1.0 – 100ul (1ul incr.) – reagent volume: 1.0 -600ul (1ul incr.) – high value sample may be diluted automatically or on request Reactionplate: – 90 high quality cuvettes – automatic cuvette washing – cuvettes Q.C. continuously computer controlled OPTIONS: – ISE model – Benchtop model – Positive barcode reader -Special language software – Barcode system DIMENSION:130cm*100cm*85cm
Reviews
There are no reviews yet.